GE Versamed Features
The Ge Versamed is a compact Ventilator that is controlled by a microprocessor. It can offer the same amount of ventilation as some of the bigger ventilators on the market – just in a smaller, portable design. The?iVent 201? is a non-invasive ventilator that runs off of a turbine. With features like NIV, Adaptive flow, and Adaptive I-Time caregivers can maximize the patient’s comfort and make it easier to breathe using the machine. Additional features of the event include 72 hours of trending of setting and martinet monitoring, a Programmable nebulizer, built-in SPo2 monitoring, and oxygen blenders. The GE iVent can be used on patients ranging from pediatrics to adults.
- Adaptive Bi-Level (NIV).
- Preset parameters by patient weight.
- Apnea backup ventilation.
- Adjustable rise time.
- Easy Exhale.
- Sigh, breathe.
- 100% O2 suction.
GE Versamed? Specifications
Dimensions
- Height: 13″ (33 cm)
- Width: 9.5″ (24 cm)
- Depth: 10.3″ (26 cm)
- Weight: Up to 27.8 lb (12.6 kg)
Ventilation Modes
- Assist/Control (A/C)
- Volume Controlled (A/C)
- Pressure Controlled (A/C)
- Synchronized Intermittent Mandatory Ventilation (SIMV)
- Volume Controlled SIMV (Vctrl)
- Pressure Controlled SIMV (Pctrl)
- Continuous Positive Airway Pressure (CPAP) with Pressure Support Ventilation (PSV)
O2 Supply Specifications
- High Pressure: Pressure Range 40 ?75psi (2.8-5.1 bar). Warning: The exact oxygen pressure which the iVent 201 uses depends on the version you have. Check the back label for the pressure levels of your unit.
- Low Pressure: Maximum flow 15L/min or 0.5 psi
Ventilation Performance and Parameters
- Respiratory Rate: 1 ? 12 ? 1, 12 ? 80 ? 2 bpm
- Tidal Volume: 50-2000 mL
- Accuracy of Tidal Volume Delivery: ?10% or ? 10 mL whichever is greater
- Accuracy of Exhale Tidal Volume Measurement: ?15% above 100 mL from actual reading or ? 10 cc below 100 mL
- Inspiratory Pressure Limit: 5 to 80? 5 cmH2O
- Inspiratory Time: Adaptive Time or 0.2 to 3 ? 10% sec
- Peak Flow (PIF): Adaptive Flow or to 120? 10% L/min Spontaneous to 180? 10% L/min
- Oxygen Mix (FiO2 ): 21% to 100% ?10% of the set value or? 3% absolute, whichever is greater
- PEEP: 0 to 40? 1 cmH2O or? 10%, whichever is greater
- Trigger Sensitivity: 1 to 20 L/min Flow, Off – 0.5 to? 20 cmH2O Pressure, Off
- PSV: 0 to 60 ?10% cmH2O
- Positive Pressure Relief Valve: 80 cmH2O
- Pneumatic Nebulizer: 5 to 240 minutes
- Controlled Pressure: 5 to 80 cmH2O ?5 cmH2O
- FiO2 at Power up: 21%, 40%, 60%, 100% (configurable)
- Configurable Humidifier Setting (Purging Cycle): HME (10 minutes), Heated (1 minute), Off
Power Supply
- External AC: 100 to 240 V, 50 to 60 Hz, Max 1.6 A
- External DC: 12 to 15 V, Max 8.5 A
- Internal battery: Sealed lead-acid 12 V (7.8 Ah) (rechargeable)
- Operating time: Depending on ventilator settings and impedance. standard internal up to 2 hours, optional extended internal up to 4 hours, and optional external up to 8 hours.
Oxygen Supply
- High-pressure supply: 40 to 75 psi (2.8 to 5.1 bar)
- Low pressure: Max 15 L/min or 0.5 psi
Environmental
- Operating temperature: 0 to 50? C/32 to 120? F
- Storage temperature: ?15 to 70? C/?4 to 140? F
- Relative humidity: 15 to 95% at 30? C/85? F
- Water/dust resistance: IP54 (splashproof)
- Atmospheric pressure: 430 to 825 mm Hg (15,000 feet)
- Vibration:
- IEC 68-2-6 and IEC 68-2-34
- MIL-STD-810E
- Shock: Total external sound level: 40 to 45 dBa at one meter
- IEC 68-2-27 (100 g)
- MIL-STD-810E
Ventilation Performance and Controlled Parameters
- Respiratory rate: 1 to 80 bpm
- Tidal volume: 50 to 2,000 ml
- Inspiratory pressure limit: 5 to 80 cm H2O
- Inspiratory time: Adaptive I-Time* or 0.2 to 3 sec
- Esens: 10-90%
- Rise time: Mid/high/max/auto
- Peak flow: Adaptive Flow* or 1 to 120 L/min; Spontaneous up to 180 L/min
- Oxygen mix (FiO2): 21 to 100%
- PEEP: 0 to 40 cm H2O
- Trigger sensitivity: 1 to 20 L/min flow sensitivity? 0.5, to 20 cm H2O pressure sensitivity
- PSV: 0 to 60 cm H2O
- Positive pressure relief valve: 80 cm H2O
- Synchronized nebulizer: 5 to 240 min
Monitoring and Displayed Parameters
- Airway pressure (analog bar graph & numerical)
- Total breath rate
- I:E ratio
- Exhaled tidal volume
- Exhaled minute volume
- Peak flow
- Inspiratory time
- Electrical power source (AC/external DC/internal DC)
- Battery level
- Pressure, flow, and volume Waveforms software package: Trending of monitored data (72 hours)
- Real-time pressure and flow waveforms
- Waveform history brows
- Respiratory mechanics (C, R, MAP, RR/Vt)SpO2
- Pressure, flow, and volume loops
Reviews